China
NADE
AC 220V (-5%~+10%), 50 Hz
32 ℃~160 ℃
0.1 ℃
adjusted to 5.0±0.5 ℃/ min automatically after three minutes
600 W
1000 ml
≤ 85%
≤ 700 W
2 or 4 samples
400mm*300mm*490mm
Product Description
The instrument is designed and made as per GB/T4507 and ASTM D36 Test Methods for Softening Point of Petroleum Asphalt of national standard of People’s Republic of China , T0608 Asphalt Softening Point Test (Ring-and-Ball Apparatus) in the Industrial Standard of People’s Republic of China JTJ052 Specification and Test Methods of Bitumen and Bituminous Mixture for Highway Engineering. It is suitable to determine softening point of petroleum asphalt, coal pitch, liquid petroleum asphalt, and various asphalts. It can be widely used for asphalt manufacturing companies, highway and bridge construction companies, relative colleges and universities, and scientific research institutes.
1. This instrument adopts microcomputer to control the test. Automatically stir, linear heating, automatically get the result and automatically print the test result.
2. The beaker adopts high temperature resisting glass material. The sizes of steel ring and ball can meet test standards.
3. Small desktop structure. LCD display. Ligh touch panel. It can determine four samples at a time. Easy to use and results reliable.
Model | SYD-2806H |
Power supply | AC 220V (-5%~+10%), 50 Hz |
Measurement range | 32 ℃~160 ℃ |
(1) Heating medium | 5℃ to 80℃ distilled water(softening point under 80 ℃) |
(2) Heating medium | 32 ℃ to 160℃ glycerin (softening point over 80 ℃) |
Temperature resolution | 0.1 ℃ |
Heating rate | It will be adjusted to 5.0±0.5 ℃/ min automatically after three minutes |
Stirrer | The electromagnetic stirring speed can be adjusted continuously |
Heating power | 600 W |
Beaker cubage | 1000 ml |
Test result | LCD screen shows results and printer to print |
Computer communication interface | RS232C communication interface |
Ambient temperature | The temperature shall be lower than 35 ℃ and keep stable. There shall not be any airflow. |
Relative humidity | ≤ 85% |
Total power consumption | ≤ 700 W |
Test sample | 2 or 4 samples |
Outline dimension | 400mm*300mm*490mm |
Notes:
1. The reagent and equipments for installing the instrument are supplied by end-users themselves.
2. The softening point thermometer needs to be purchased by the end user. Parameters: 30 ℃ ~ 180 ℃; scale is 0.5 ℃.
3. The usage,management and saving of poisonous and harmful reagent,pls operate according to the relative principle.
4. PLS read the operation manual and industry standard before using the instrument.
Other models:
content is empty!