| Availability: | |
|---|---|
PC950
NADE
DC9V
1
1 year
China
Product Description



1. Comes with TestBench that incorporates a flexible electrode holder, an intelligent stirrer, and buffer organizers altogether, streamlining the process of pH measurement and calibration.
2. Intuitive operating navigation with natural language and graphics
3. Multi-language operating system (English, German, Spanish, Mandarin)
4. Automatic calibration, automatic temperature compensation, self-diagnosis.
5. Manual data logger with 200 groups of storage
1. The 2401T-F glass conductivity/temp.electrode designed for research-level conductivity measurement from 0 to 200mS/cm at ±1% F.S accuracy
2. 1 to 4 points automatic calibration with calibration instruction and self-diagnosis
3. Single tap to switch from conductivity to TDS and salinity
4. Automatically recognizes conductivity standard solution (84µS/cm, 1413µS/cm, 12.88mS/cm, 111.8mS/cm).
5. Standard BNC conductivity connector, RCA connector for a temperature probe, NTC 30KΩ thermistor
Specification
Technical Specifications | ||
Cond. |
Measuring Range | Conductivity:0~200 mS/cm, divided into five ranges, automatically switch measuring range. |
Resolution | 0.01/0.1/1μS/cm 0.01/0.1 mS/cm | |
Accuracy | ±1.0% FS ±1 digit | |
Temperature Compensation Range | (0 to 50)˚C (Auto. or Manual) | |
Cell constant | 0.1 / 1 / 10 cm-1 | |
Temp. | Measuring Range | 0~100℃ |
Resolution | 0.1℃ | |
Accuracy | ±0.5℃±1 digit | |
Other Specifications | |
Data Storage | 100 groups |
Storage Content | Number, measuring value and temperature value |
Power | Input: 120V AC 60Hz; |
Dimension & Weight | Meter:(360*165*400 )mm / 1.25kg |
Connector | BNC for conductivity; RCA for temperature |
What's in the box? | EC950 Meter, 2401T-F conductivity/temp. electrode, 9V power adapter, 606 combination test bench ( with stirrer ), 50ml of conductivity standards (84µS, 1413µS, 12.88mS), instruction manual |
content is empty!