China
NADE
AC(220±10%)V, 50Hz
0 penetration~600 penetrations
0.1 penetration
5s or 60s, can be setted at will,bias within ±0.1s
200 W
25±0.1 ℃
High accuracy digital temperature controller
Hard glass bath
2.5g±0.05g
280mm*350mm*700mm(L*W*H)
Product Description
This instrument is designed and made as per T0604-2011 Penetration Test for Bituminous Materials in JTG E20-2011 Specification and Test Methods of Bitumen and Bituminous Mixture for Highway Engineering and ASTM D5 Standard Test Method for Penetration of Bituminous Materials. It is used to determine the penetration of pavement petroleum asphalt, modified asphalt, liquid petroleum asphalt and emulsified asphalt. It is also suitable to test solid particle, powder, colloid and raw-food materials such as cheese, glycine, butter, cream and leavening. It can be widely used in food industry, highway engineering and other industrial fields.
1. The needle holder is designed novelly. It can be moved vertically and adjusted roughly and finely without measurable frication.
2. It adopts novel heating design. The heating and temperature measuring appliance is separated from water bath. Easy to operate.
3. Equipped with automatic temperature controller and high accuracy temperature sensor, the temperature can be set conveniently and control precision is high.
4. It adopts newly designed digital meter. The measuring data is stable and reliable.
Model | SYD-2801E1 |
Measurement range | (0~600) penetration scale |
Resolution | 0.01mm |
Timing controller | 5s or 60s, can be setted at will,bias within ±0.1s |
Heating power | 200W |
Temperature control accuracy | 25℃±0.1℃ (Note: The ambient temperature shall not exceed 20℃) |
Temperature control mode | High accuracy digital temperature controller |
Constant temperature bath | Hard glass bath |
Stirring | Automatically stirred by magnetic stirrer |
Standard needle | 2.5g±0.05g |
Needle holder | It can be adjusted roughly and finely without measurable frication to make the tip aim the sample conveniently. |
Power supply | AC(220±10%)V, 50Hz |
Dimension | 280mm×350mm×700mm |
Optional accessories
1. Grease masher : It is used in the tests which determine the cone penetration of lubricating grease (or petrolatum)
2. Standard cone: 102.5g±0.05g
3. Other cones:1/2 scale cone,1/4 scale cone
Notes:
1. The reagent and equipments for installing the instrument are supplied by end-users themselves.
2. The Mercury thermometer needs to be purchased by the end user. Parameter:0~50 ℃, scale division is 0.1 ℃).
3. The usage,management and saving of poisonous and harmful reagent,pls operate according to the relative principle.
4. PLS read the operation manual and industry standard before using the instrument.
Other models:
content is empty!