EC400F
AC Adapter, 100-240V AC input, DC9V output
1
1 year
China
Product Description
● High resolution LCD display, 5.7 inches.
● Multi-reading feature allows auto-read, timed-read and continuous-read.
● Automatic/Manual temperature compensation ensures accurate results.
● Auto-hold feature senses and locks the measurement endpoint.
● Data Storage 500 sets (GLP-compliant).
● Support for USB or RS-232 communication.
● Reset feature automatically resumes all settings back to factory default options.
● IP54 waterproof.
● 1-3 points calibration with Standard Recognition.
● Settable parameters, including cell constant, temperature compensation coefficient and TDS factor.
● Temperature compensation type (none, linear, pure water).
Specification
Model | EC400F | |
Parameters | EC/Temp.(Resistivity/TDS/Sal.) | |
Conductivity | Range | 0.000 μ S/cm to 1000 mS/cm |
Resolution | 0.001 μ S/cm minimum, various with range selection | |
Accuracy | ± 0.5% FS | |
Reference Temperature | 20, 25 ºC | |
Calibration Points | Up to 3 | |
Standard Recognition | 84 μ S/cm, 1413 μ S/cm, 12.88mS/cm | |
Resistivity | Range | 5.00 Ω ·cm~100.00 M Ω ·cm |
Resolution | 0.01 Ω ·cm minimum | |
Accuracy | ± 0.5% FS | |
TDS | Range | 0.00mg/L~1000g/L |
Resolution | 0.001mg/L minimum, various with range selection | |
Accuracy | ± 0.5% FS | |
Salinity | Range | (0.00~8.00)% |
Resolution | 0.01% | |
Accuracy | ± 0. 1% | |
Temperature | Range | - 10 to 135 ºC, 14 to 275 ºF |
Unit | ºC, ºF | |
Resolution | 0.1 | |
Accuracy | ± 0. 1 | |
Measurement | Reading Mode | Auto Read (Fast, Medium, Slow), Timed, Continuous |
Reading Prompts | Reading, Stable, Locked | |
Temp. Compensation | ATC, MTC | |
Data management | Data Storage | 500 results each |
GLP Features | Yes | |
Conductivity with Temp. Probe | 5-pin aviation connector | |
Outputs | USB | USB 2.0 flash memory device, PC |
RS 232 | printer | |
Display options | Backlight | Yes |
Auto Shutdown | 1~60 min, off | |
IP Rating | IP54 | |
Date and Time | Yes | |
General | Power | AC Adapter, 100-240V AC input, DC9V output |
Dimensions | 242 × 195 × 68 mm | |
Weight | 900g (1.98 lb) |
Configuration
DJS-1VTC EC Electrode
1413μS/cm Conductivity Standards Solution 1 Bottle (50ml/bottle)
REX-5 Multifunction Electrode Stand
PHSJ-308F