Availability: | |
---|---|
WVTR-I1
NADE
AC 220V 50Hz
1
1 year
China
Product Description
It is designed and manufactured based on the infrared sensor method, used to measure the water vapor transmission rate of barrier materials with high and medium moisture barrier properties with a wide testing range and high testing efficiency. WVTR-I1 Water Vapor Permeability Tester is applicable to determination of water vapor permeability of plastic films, laminated film, high barrier materials, back sheet, sheeting, paper, sheet metal, packages and other relative packaging materials in food, pharmaceutical, medical apparatus, consumer goods, photovoltaic and electronic industries, etc. It can also be extended to test the water vapor permeability test of bottles, bags, tubes and other containers.
Test Principle
The test specimen is mounted in the diffusion cell, which is divided into a dry chamber and a controlled-humidity chamber. The dry side of the specimen is swept by a flow of dry nitrogen, and the water vapor permeating through the specimen from the controlled-humidity chamber is carried by dry nitrogen to the infrared sensor where proportion electrical signals will be generated. The water vapor transmission rate is obtained by analyzing and calculating the electrical signals.
Feature
Application
Basic Application | Film | Water vapor transmission rate of plastic films, paper-plastic composite films, coextruded films, aluminum foils, aluminum composite films, glass fiber aluminum foil composite films and many others. |
Sheet | Water vapor transmission rate of PP, PVC and PVDC sheets, metal foils, rubber pads, silicon wafers and other sheeting materials. | |
Paper, paper board and composite materials. | Water vapor transmission rate of cigarette aluminized paper, paper aluminum-plastic composite sheet and other paper and cardboard | |
Packages | Water vapor transmission rate of plastic, rubber, paper, paper-plastic composite, glass and metal packages, e.g. plastic bottles, pouches, coated paper cartons, vacuum bags, metal three-piece cans, plastic packages for cosmetics, soft tubes for tooth paste, jelly and yogurt cups. | |
Extended Application | Package closure | Water vapor transmission rate of all types of package closure |
LCD display | Water vapor transmission rate of LCD display and its relevant sheet. | |
Solar Back-sheets | Water vapor transmission rate of solar Back-sheets and its relevant sheet. | |
Pipe | Water vapor transmission rate of PPR and other types of pipes | |
Blister Packs | Water vapor transmission rate of whole blister packs | |
Aseptic protective film, medical plaster patch | Water vapor transmission rate of aseptic protective film, medical plaster patch | |
Battery Plastic Shell | Water vapor transmission rate of battery plastic cell |
Specification
Model | WVTR-I1 |
Test Range | Film:0.001~ 100 g/m2·24h (Normal) |
Container:0.00001~ 0.25 g/pkg·24h (Optional) | |
Resolution | 0.001 g/m2 ·24h |
Specimen Area | 50cm2 |
Specimen Thickness | ≤3 mm (Accessories needed for other thickness) |
Test Mode | Single chamber |
Test Temperature | 5℃~ 95℃ |
Temperature Accuracy | ±0.1℃ |
Test Humidity | 0% RH,35%~95%RH ,100%RH (Auto-control) |
Humidity Accuracy | ±1%RH |
Carrier Gas | 99.999% High-purity Nitrogen (outside of supply scope) |
Carrier Gas Flow | 0~200 mL/min(Auto-control) |
Gas Supply Pressure | ≥0.2MPa |
Port Size | 1/8" metal tube |
Dimensions | 585 mm (L)×640 mm (W)×380mm (H) |
Power Supply | AC 220V 50Hz |
Net Weight | 50Kg |
content is empty!