NADE JK9870A Fully Automatic Kjeldahl Analyzer

Analysis range: 0.1—240 mg N
Precision (RSD): ≤0.5%
Recovery rate: 99-101%
Touch screen: 10-inch color LCD touch screen
Availability:
  • JK9870A

  • AC220V/50Hz

  • 1

  • 1 year

  • China

Product Description

JK9870A

The Kjeldahl method is a classic technique for determining nitrogen.

It is widely used for the measurement of nitrogen in soil, food, livestock, agricultural products, feed, and other nitrogen-containing compounds. The process for sample analysis using this method involves three main steps: sample digestion, distillation and separation, and titration analysis.

Our company is one of the creators of the national standard GB/T 33862-2017 for "Fully (Semi) Automated Kjeldahl Nitrogen Analyzer." Therefore, the Kjeldahl analyzer series products we develop and manufacture comply with both the GB standard and relevant international standards.


Technical Specification

Analysis range: 0.1—240 mg N

Precision (RSD): ≤0.5%

Recovery rate: 99-101%

Minimum titration volume: 0.2 μL/step

Titration speed: 0.05—1.0 ml/s, adjustable

Distillation time: 10—9990 seconds, adjustable

Sample analysis time: 4-8 minutes (cooling water at 18°C)

Titrant concentration range: 0.01—5 mol/L

Titrant concentration input method: Manual input/instrument internal standard

Titration mode: Standard/Simultaneous distillation and titration

Titration cup capacity: 300 ml

Touch screen: 10-inch color LCD touch screen

Data storage capacity: 1 million sets of data

Printer: 5.7 cm thermal printer with automatic paper cutting

Communication interface: RS232/Ethernet/electronic balance/cooling water/reagent bucket level

Bluetooth communication, data transmission, remote monitoring

Digestion tube waste discharge mode: Manual/automatic

Steam flow regulation: 1%—100%

Safe alkali addition mode: 0—99 seconds

Automatic shutdown time: 60 minutes

Working voltage: AC220V/50Hz

Heating power: 2000W

Main unit dimensions: 500 mm (L) x 460 mm (W) x 710 mm (H)

Cooling system temperature range: 5°C-20°C

Cooling tank capacity: 6L

Circulation pump flow rate: 10L/min

Lift: 10 meters

Working voltage: AC220V/50Hz

Power: 850W