| Availability: | |
|---|---|
KFT-41C
NADE
1
1 year
China
Features
Features
● Color High contrast LCD touchscreen, 7.0 inches.
● Coulometric KF Titration.
● Support oven sampling method, to meet the gas, solid, liquid in the constant moisture, trace moisture measurement.
Coulometric Karl Fischer Titration
● Support Auto Titration Method, Oven Titration Method, Oven Blank, and Coefficient correction Method.
● The solvent manager adopts anti-leakage liquid and anti-pourdesign of waste liquid bottle, supports to drain solution andfeed solvent, ensure safe handling of Karl Fischer chemicals.
Data management
● Auto/manual drift adjust ensures accurate results.
● Selectable units including μg, mg, %, ppm, mg/L and μg/mL.
● User password management.
● Store up to 200 titration data sets (GLP-compliant).
● Support for RS-232 communication for print and the print format can be alternate.
● Support for USB communication for PC software.
● Support for U disk for data export with CSV format.
● It needs GMP software to implement GMP management functions.
Specification
Model | KFT-41C | |
Principle | Karl Fischer Coulometric Titration | |
Water | Range | 10.0μg~200mg |
Resolution | 0.1μg | |
Unit | μg,mg,%,ppm,μg/mL,mg/L | |
Repeatability | ≤1% | |
mV Range | (0~2000)mV | |
mV Resolution | 0.1mV | |
Polarization Current Accuracy | ±3% | |
Polarization Current Fluctuation | ±2.5%/30min | |
Working Current Accuracy | ±0.5% | |
Working Current Fluctuation | ±0.2%/10min | |
Power Supply | (100~240)VAC,(47~63)Hz | |
Dimensions | 240×370×270mm | |
Net Weight | About 4kg | |
● Include
Solvent manager
Coulometric vessel
Measuring electrode
Generator electrode
content is empty!