You are here: Home / Products / Water quality analyzer/tester / others / Bante322 Water Hardness Meter

loading

Share to:
facebook sharing button
twitter sharing button
line sharing button
wechat sharing button
linkedin sharing button
pinterest sharing button
sharethis sharing button

Bante322 Water Hardness Meter

Range (Concentration):0.05 to 200mmol/L
Range (CaCO3):0 to 19999mg/L
Range (CaO):0 to 11220mg/L
Range (Ca2+):0 to 8020mg/L
Availability:
  • Bante322

  • 1

  • 1 year

  • China

Product Description

Bante322 Portable Water Hardness Meter

1.jpg

2.jpg            3.jpg

Professional portable water hardness meter comes with a solid-state ion selective electrode, the meter is suitable for outdoor applications, accuracy: 1% F.S.


Feature

• Professional portable water hardness meter is equipped with a backlit LCD display.
• 2 to 5 points calibration from low to high concentrations.
• Selectable water hardness units - mmol/L, mg/L, german degree (°dH), english degree (°e) and french degree (°fH).
• Automatic temperature compensation ensures accurate readings over the entire range.
• Auto-Read function senses and locks the measurement endpoint.
4.png
• Calibration due alarm prompts user to calibrate the meter regularly.
• Auto-Power Off effectively conserves battery life.
• Setup menu allows setting the number of calibration points, stability criteria, etc.
• Reset function automatically resumes all settings back to the factory defaults.
• Expanded memory stores or recalls up to 500 data sets.
• Multi-mode power scheme (batteries, power adapter and USB port) ensures that using the meter smoothly.

Specification

Model


Bante322

  Water Hardness



  Range (Concentration)


0.05 to 200mmol/L

  Range (German Degree)


0 to 1122°dH

  Range (English Degree)


0 to 1404°e

  Range (French Degree)


0 to 2000°fH

  Range (CaCO3)


0 to 19999mg/L

  Range (CaO)


0 to 11220mg/L

  Range (Boiler)


0 to 400mmol/L

  Range (Ca2+)


0 to 8020mg/L

  Resolution


0.001, 0.01, 0.1, 1

  Accuracy


±1% F.S.

  Calibration Points


2 to 5 points

  Calibration Solutions


0.01, 0.1, 1, 10, 100mmol/L

  Temperature



  Range


0 to 105°C

  Resolution


0.1°C

  Accuracy


±0.5°C

  Offset Calibration


1 point

  Calibration Range


Reading ±10°C

  General Specifications



  Temperature Compensation


0 to 50°C, manual or automatic

  Stability Criteria


Low or high

  Calibration Due Alarm


1 to 31 days or off

  Auto-Power Off


10, 20 or 30 minutes after last key pressed

  Memory


Stores up to 500 data sets

  Output


USB communication interface

  Connector


BNC, 3.5mm jack socket

  Display


Custom LCD (80×60mm)

  Power


3×1.5V AA batteries or DC5V power adapter

  Battery Life


Approximately 150 hours (turn off the backlight)

  Dimensions


170(L)×85(W)×30(H)mm

  Weight


300g

Bante322.jpg


Real photos show

IMG_0452.JPGIMG_0458.JPGIMG_0449.JPG

Previous: 
Next: 

googleTranslate