Home/Products/Nade XT-9912 Intelligent Microwave Digestion/Extraction System
Nade XT-9912 Intelligent Microwave Digestion/Extraction System
Nade XT-9912 Intelligent Microwave Digestion/Extraction System
Size: XT-9912
Note:
Category: Digestion/extractor
 
Detailed Product Description
 
1.Intelligent Microwave Digestion
2.high efficiency of sample treating
3.High pressure exterior vessel
4Overheating protection
Intelligent Microwave Digestion Parameter
 
1.Microwave output power and frequency: 0-1300W, stepless adjustment; 2450 MHz
2.Microwave resonance radiation cavity: 65L, internally installed corrosion resistant internally clapboard
3.Multi-vessel pressure monitoring: 0.1~5MPa (precision:±0.01MPa); multiple non-contact pressure sensors, real-time monitoring, display and control the true pressure of 12 digestion vessels
4.Multi-vessel temperature monitoring: 0-250°C(precision:±1°C); multiple non-contact temperature sensors, real-time monitoring, display and control the true temperature of 12 digestion vessels
5.Number of digestion sample: simultaneously digest 12 samples of same or different interstitial substance contents. The user may freely digest or extract 1-12 samples
 
Multiple safety measures:
      1.High pressure exterior vessel: PSF exterior vessel with PPS   liner,      maximum pressure endurance is 6MPa (900 psi).
      2.Overheating protection: when the case or the exterior vessel is overheated, instrument will warn and suspend the microwave radiation
      3.Abnormal noise detection: when abnormal noise is made in the furnace chamber, instrument will warn and cut off microwave automatically.
      4.Abnormal pressure warning: a unique device of ‘optical fiber non-contact displacement sensor’, real-time monitor and control the pressure in the digestion vessel, display every vessels’ pressure change by screen. Independently monitor each vessel’s reaction situation, avoid overpressure caused by single sample’s too violent reaction. Pressure auto matching control and rise speed limitation, when the pressure is much too high or the rise speed is too fast, instrument warns and cuts off the microwave output.
      5.High strength bursting framework and design of dual protective door, effectively assure the safety of operator.
      6.Dual pressure reduction by inner vessel’s air bleed hole and bursting film: when the gas pressure in the digestion vessel is too large, gas will leak from the air bleed hole; when the pressure is larger than the limit value and the hole can’t leak the air in time, the skirt border of the sealing bowl will broken, then leak gas and reduce pressure.
       7.Double security door